Shanghai vacuum drying box failure solution

In the process of using the vacuum drying oven, more or less will encounter large and small faults, do you understand the solution to the vacuum drying oven in Shanghai ? Below, Dongsheng Instrument introduces the solution to the problem of Shanghai vacuum drying oven :

Fault phenomenon

Reason speculation

Approach

Power on without power

External power outlet has no electricity

Check if the line is tripped, is the socket good?

The power cord plug is not plugged in or disconnected

Reinsert or repair the power cord

The fuse is not installed or broken

Check that there is no short circuit in the vacuum box, and the fuse tube is shorted (the instrument power transformer is short-circuited, the heater is short-circuited, and the ground short-circuit causes the fuse to blow).

Meter PV screen display

" □ □ □ □ ”

Temperature sensor Pt100 is bad

Check Pt100 and replace ( 0 °C for 100 Ω , 0.3 Ω / °C )

Temperature sensor wiring off

Rewiring

Meter measurement range is wrong

Reset

Not warming up

Low setting temperature

Set temperature SV ≥ RT+10 °C

RT is (ambient temperature)

Instrument output circuit wiring is off

Rewiring

The temperature controller has no output signal or bad or thyristor

Exchange

Electric heater is bad (open circuit, short circuit)

Exchange

Not warming up

Enable timing or incorrect settings

ST=0 or ST= (heating + constant temperature) time

Temperature out of control or set temperature and measured temperature error is out of tolerance to generate static difference, overshoot

Temperature controller output is out of control

3041 or BTA is bad, change

Does not meet the conditions of use

SV ≥ RT+10 °C

Poor contact of Pt sensor

Clear contact resistance

The parameter setting is incorrect.

Reset the relevant parameter example P, etc.

Measuring temperature and

The actual temperature in the box is out of tolerance

No vacuum

Vacuum degree

Measuring the mercury thermometer is not on the shelf

Reposition

Meter or parameter change

Re-correct SC and HL parameters

Can't vacuum

Vacuum pump selection model specifications are wrong

The pumping rate should be no less than 2 liters / sec

Various connecting pipes (heads) are loose and the inner diameter is too fine

Re-select the appropriate inner diameter connecting tube and fasten the joint

Vacuum gauge is bad

replace

The door is not closed

Adjust the door distance

Door seal rubber aging loses elasticity

Replace the door seal

The position of the purge valve and vacuum valve is incorrect.

Adjust position

Can't vacuum

Solenoid valve is bad ( type 6090, type 6210 )

Replace the solenoid valve

Air leak

( vacuum within 24 hours

Down by the -0.1Mpa

Greater than -0.192Mpa )

Various connecting pipes leaking

Replace after inspection

Except for 6050 , 6050B , 6051 , 6053 type, the heater " O " type seal is deformed and causes no compression.

Tighten the heater base (back of the liner) with an Allen key or replace the “ O ” seal

The bleed valve is not in the correct position

Place the correct position

Vacuum valve leak

Exchange

Solenoid valve is not leaking tightly (applicable to 6090 , 6210 )

Exchange